RESmart®Auto CPAP

Ergonomic Design

  • Integrated heated humidifier
  • DC 24V powered and infrared controlled humidifier for patient safety and comfort
  • Delay-off feature protects the device from humidity hazard
  • Backlit LCD display for operation in dark
  • Light user buttons with lockup capability
  • Auto adjustment of therapy pressure within set range
  • Personalized sensitivity setting
  • 4-20 cmH2O therapy pressure
  • Real time audible alert for tubing/mask leak
  • RESlex® exhalation relief with three comfort settings
  • Remote compliance data access by phone with iCode®
  • SD Card Integration with cloud based iCode Connect
  • Embedded memory stores last night’s full raw data, 365 nights of user data and more than 30 years of therapy records
  • Last night’s compliance-therapy snapshot to assist patient with compliance
  • Automatic leakage and altitude compensation
  • Inspiration trigger for automatic start-up
  • Quiet operation

Download (PDF, 331KB)

Dimensions: 8.66 × 7.6 × 4.4” (220 x 194 x 112mm), 12.3” (313 x 194 x 112mm) (with Integrated Heated Humidifier)
Weight: 4.9 lbs. (2.2 kg), 5.3 lbs. (2.4 kg) (with Integrated Heated Humidifier)

Environmental parameters
Operating Transport and storage
Temperature 41˚F to 86˚F (5 to 30˚C) -4˚F to 131˚F (-20 to 55˚C)
Humidity: ≤ 80% non-condensing ≤ 93% non-condensing
Atmospheric pressure: 860 to 1060 cmH2O 500 to 1060 cmH2O

Mode of Operation:
Continuous
Work Mode: CPAP, Auto SD Card Capacity: 2G bytes AC Power Consumption: 100 – 240V AC, 50/60Hz , 1.0A max
Type of Protection Against Electric Shock: Class II Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part Degree of Protection Against Ingress of Water: IPX1 –Drip-Proof, Vertical Pressure Range: 4 to 20 cmH2O * (in 0.5 cmH2O increments) Pressure Stability: 4 to 20 cmH2O (±1 cmH2O)
Altitude setting: Manual setting:, level 0, 1, 2